Figure 6-26 Two point calibration flat field image analysis
Selected bad pixels will be shown as below.
Green and red pixels are from the previous step.
Yellow represents pixels with higher PRNU.
Purple represents pixels with higher temporal noise.
Figure 6-27 Selected bad pixel from flat field image analysis
Select “Finish” and save the file.
*** Additional tip
– It is still possible to correct bad pixels not detected by the calibration
wizard, during the two point calibration procedure, by manually editing the bad pixel map.
Refer to Xeneth user manual for more information (Ref 1).