Figure 5-1 Show camera property documentation
Image acquisition
Gain settings
There are 4 analogue gain options on the Manx camera. It can be selected by the user.
Sensor Gain Mode
Full well (electrons)
1
35M
5
8.1M
20
1.6M
100
290K
Table 5-1 Analogue gain settings
Figure 5-2 Gain selection on Xeneth
* x500 is an additional gain setting which is not tested
A simplified maximum Signal to Noise equation can be written as below. (The formula
neglects noise contribution from dark current and camera read noise)
𝑆𝑁𝑅
max
=
𝑆𝑖𝑔𝑛𝑎𝑙
max
𝑠𝑞𝑟𝑡 (𝑆𝑖𝑔𝑛𝑎𝑙
max
)
= 𝑠𝑞𝑟𝑡 (𝑆𝑖𝑔𝑛𝑎𝑙
max
) = 𝑠𝑞𝑟𝑡 (𝐹𝑢𝑙𝑙 𝑊𝑒𝑙𝑙)
Equation 1 Max signal to noise ratio
Based on Equation 1, it is better to use the camera with lower gain factor, if there is a huge
influx of incoming signal.