The next step requires a uniform dark image. It corrects for Fixed Pattern Noise (FPN) which
is caused by differences in individual pixel offset.
We recommend using a threshold of 350% for both FPN and temporal noise.
Select “Perform capture and analysis” to proceed.
Figure 6-21 Two point calibration