When the exposure time is less than acquisition line time, the actual line time equals to the
acquisition line time. Depending on the situation, increasing the actual line time may help to
improve the image aspect ratio.
Refer to the frame rate calculator if necessary. (Ref 6)
Acquisition frame time
The frame rate is affected by the line rate and minimum frame time defined by the user.
Each frame comprises of a user defined number of lines. See section 5.1.6 to define the
number of lines. Frame time refer to the time taken to capture N number of lines. The
minimum time between 2 frames can be defined by the user by setting the Acquisition frame
time
Figure 5-5 Acquisition frame time setting on Xeneth
Refer to the frame rate calculator if necessary. (Ref 6)
Acquisition mode
There are 2 acquisition modes describe as below:
Acquisition control
Acquisition
Mode
Multi frame Stop acquisition command is automatically activated when
Acquisition Frame count equals Frame count captured.
Continuous This is the default mode. Frames are captured continuously
until a stop acquisition command is given by the user