Line and frame triggers can be configured as shown below:
Trigger in source configurations
Frame trigger
Free
running
External
trigger 1
External
trigger 2
CoaXpress
Line
trigger
Free running
x
x
x
x
External
trigger 1
x
x
x
External
trigger 2
x
x
x
CoaXpress
x
x
x
Table 5-7 Trigger in source configurations
Lines(frames) can be generated in the following modes:
Line(frame) trigger in polarity configuration
Trigger in
sensitivity
Trigger in
polarity
Edge
Low-falling
Lines(frames) are generated when the trigger in voltage
polarity is falling
High-raising
Lines(frames) are generated when the trigger in voltage
polarity is raising
Level
Low-falling
Lines(frames) are continuously generated when the
trigger in voltage polarity is low
High-raising
Lines(frames) are continuously generated when the
trigger in voltage polarity is high
Table 5-8 Line(frame) trigger in polarity configuration
Skipping of triggers and trigger in delay can be adjusted. This is how it is shown on Xeneth
user interface.