Table 5-3 Footer contents of the Manx
Tec control
The built-in thermo-electric cooler (TEC) can be regulated via Xeneth software. It is
recommended to keep the target temperature at 20degC in room temperature conditions.
The purpose of the TEC is to maintain a stable sensor temperature so that the dark current
and hence the dark noise is kept constant.
Figure 5-10 Tec regulation settings on Xeneth
Be careful not to overheat the camera. It is advisable to use the camera with an additional
heat sink and/or fan.
10
Do not use
Not used, value = 0
11
Do not use
Not used, value = 0
12
Do not use
Not used, value = 0
13
Do not use
Not used, value = 0
14
Do not use
Not used, value = 0
15
Do not use
Not used, value = 0
16
Do not use
Not used, value = 0
Maintain sensor temperature below 50
C to avoid risk of damaging the
camera!
!