Local Operation
3-36
975-0200-01-01
To reset the power supply:
1. Turn the front panel power switch to the Off position.
2. Allow the unit to power down.
3. When the fans are off and the front panel displays are not illuminated,
turn the power switch to the On position.
4. Press and hold the OUTPUT ON/OFF and AUX ON/OFF buttons.
The front panel should display
P5U S rS
.
5. When the display reads
P5u Clr
, release the OUTPUT ON/OFF and
AUX ON/OFF buttons. After 1 second the model number will appear
on the display for 1 second.
The power supply has been reset.
Foldback delay
0.5 s
Analog programming
Off
Analog programming range
Max
| Current Share Mode | MASTer |
Off
Alarm Output Latches
0 × 87 (or all latches enabled)
Alarms Mask
0 × FF (all enabled)
Interlock
Disabled
Voltage Analog Programming
Off
Voltage APG Scale
10 V
Current Analog Programming
Off
Current APG Scale
10 V
Table 3-9
Power Supply Default Settings
Parameter
Setting