Remote Operation
5-60
975-0200-01-01
:PROG:STAT RUN
The output after the last command might be as follows:
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Repeating an Auto Sequence Program:
An Auto Sequence Program can be configured to repeat any number of
times or infinitely. The default at power up is to have the repeat count
default to once. The program will repeat if the repeat count can be
changed using the
:PROGram[<channel>]:REPeat
command. If a
program is running or pause the
:PROGram[<channel>]:REPeat
command will not work.
Commands:
:PROGram[<channel>]:REPeat {?|<count>|INFinity}
Where:
<count> is the repeated count from 1 to 65 534.
INFinity is the character mnemonic for repeating without end.
For Example:
:PROG:STAR
*IDN?
*IDN?
:PROG:STOP
:PROG:REP 2
:PROG:STAT RUN
The output after the last command might be as follows:
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Important:
Execution of the program may be terminated at any time by
pressing the Esc key in the MS Windows Hyper Terminal window. The power
supply can be controlled during program execution.
Содержание XTR 100-17
Страница 2: ......
Страница 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 48: ...2 18...
Страница 86: ...Local Operation 3 38 975 0200 01 01 Figure 3 4 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 122: ...4 30...
Страница 208: ...6 20...
Страница 244: ...A 36...
Страница 250: ...B 6...
Страница 264: ...C 14...
Страница 272: ...IX 4...
Страница 273: ......