background image

 

CAEN

 

Tools for Discovery

 

n

 

Electronic Instrumentation

 

 

 

 
 

 

Rev. 0 - 

 

 
 
 

 

 

Es 

 

 

 

A1539, AG539, A154X, AG54X  Power Supply 

Boards 

Rev. 2 - 6 May 2015

 

 

 

Содержание A1539

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 0 Es A1539 AG539 A154X AG54X Power Supply Boards Rev 2 6 May 2015...

Страница 2: ...ccident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It is strongl...

Страница 3: ...safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 4 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 12 Sa...

Страница 4: ...rs 24 channel SHV versions are double width boards 10 TE all other versions are single width 5 TE All boards are provided with both current and voltage protections If over current occurs the relevant...

Страница 5: ...nels 0 500 V common to all board channels VMAX hardware accuracy 1 V VMAX software 0 100 V settable for each channel 0 500 V settable for each channel VMAX software resolution 1 V Ramp Up Down 1 50 Vo...

Страница 6: ...single width 5 TE height is 6U External connections The function and electro mechanical specifications of the external connectors are listed in the following subsections Version 24 32 channel DB25 12...

Страница 7: ...ntation A1539 AG539 A154X AG54X Power Supply Boards 7 DB25 connector pin assignment Fig 2 DB25 connector pin assignment 32 channel version INT INT interlock RTN channel return AGND earth SH connector...

Страница 8: ...s VMAX trimmer Function it allows to adjust the hardware maximum voltage VMAX common to all the channels Its value can be read out via software Shield CFG jumpers 24 32 ch floating Function see table...

Страница 9: ...s To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or d...

Страница 10: ...ameter shows the ON OFF channel status As this parameter is set ON the channel is switched on if the INTERLOCK is not active and if the channel is enabled either locally or remotely highlighted in gre...

Страница 11: ...r it is possible to monitor board parameters such as measured Temperature and HVMax and to check board status the following messages may be returned by the POWER SUPPLY SYSTEM when monitoring the boar...

Страница 12: ...is available as front panel 2mm socket connector This feature allows on detector grounding thus avoiding loops which may increase noise level FAGND and AGND may be coupled in several ways according to...

Страница 13: ...HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS 1 HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN...

Страница 14: ...including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingen...

Отзывы: