A1539, AG539, A154X, AG54X  Power Supply Boards

 

12 

 

Boards with AMP interlock connectors 

 

 

Short circuit pin 3 and 4 [the lower ones]. Leave contact open between pin 1 (+) and 2 
(-) [upper ones] 

 

Apply +5 V (3-4mA) differential signal between pin 1 (+)  and 2 (-); pin 1 is the upper 
one. Leave contact open between pin 3 and 4. 

 
The INTERLOCK LED (red) is turned off as one of the actions above is performed. 
When the channels are disabled the voltage outputs drop to zero at the maximum rate available; when 
the output disable cause is removed (see above), the channels remain OFF until the User turns them ON 
via software.. 

Grounding specifications 

The Mod. A1539 - A154X channels share a common floating return (FAGND, see External connections), 
available on the  front  panel  multipin connector, insulated from the  crate  ground (AGND,  see External 
connections), 
which is available as front panel 2mm socket connector.  This feature allows on-detector 
grounding,  thus  avoiding  loops  which  may  increase  noise  level.  FAGND  and  AGND  may  be  coupled  in 
several ways, according to environment requirements. Examples refer to 32 channel version. 

Safety Earth connection 

The connection of shield and return to Earth is fundamental for User safety.  
The connection must always be at the level of detector or power supply system.  
Return and Shield connections even if not present or performed incorrectly, due to protection circuits 
implemented  on  the  A1539  -  A154X  are  bound  to  Earth;  in  this  case  the  voltage  difference  between 
return and Earth (System),  shield and Earth is limited  to  approximately 50V.  Please note that  this is a 
status of emergency-protection, not a working one. The Connector Configurator allows to optimize the 
connection of the shield, of the return and of AGND (Earth). The best configuration must be determined 
by  the  user  upon  application,  the  optimal  connection  depends  on  many  characteristics  of  the  related 
experiment.  

Содержание A1539

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 0 Es A1539 AG539 A154X AG54X Power Supply Boards Rev 2 6 May 2015...

Страница 2: ...ccident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It is strongl...

Страница 3: ...safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 4 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 12 Sa...

Страница 4: ...rs 24 channel SHV versions are double width boards 10 TE all other versions are single width 5 TE All boards are provided with both current and voltage protections If over current occurs the relevant...

Страница 5: ...nels 0 500 V common to all board channels VMAX hardware accuracy 1 V VMAX software 0 100 V settable for each channel 0 500 V settable for each channel VMAX software resolution 1 V Ramp Up Down 1 50 Vo...

Страница 6: ...single width 5 TE height is 6U External connections The function and electro mechanical specifications of the external connectors are listed in the following subsections Version 24 32 channel DB25 12...

Страница 7: ...ntation A1539 AG539 A154X AG54X Power Supply Boards 7 DB25 connector pin assignment Fig 2 DB25 connector pin assignment 32 channel version INT INT interlock RTN channel return AGND earth SH connector...

Страница 8: ...s VMAX trimmer Function it allows to adjust the hardware maximum voltage VMAX common to all the channels Its value can be read out via software Shield CFG jumpers 24 32 ch floating Function see table...

Страница 9: ...s To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or d...

Страница 10: ...ameter shows the ON OFF channel status As this parameter is set ON the channel is switched on if the INTERLOCK is not active and if the channel is enabled either locally or remotely highlighted in gre...

Страница 11: ...r it is possible to monitor board parameters such as measured Temperature and HVMax and to check board status the following messages may be returned by the POWER SUPPLY SYSTEM when monitoring the boar...

Страница 12: ...is available as front panel 2mm socket connector This feature allows on detector grounding thus avoiding loops which may increase noise level FAGND and AGND may be coupled in several ways according to...

Страница 13: ...HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS 1 HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN...

Страница 14: ...including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingen...

Отзывы: