Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-12
975-0200-01-01
Voltage-Controlled Current APG Setup
Activating APG Current Mode
To activate APG current mode using an external voltage source:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr
is displayed on the output voltage display.
2. Turn the Rotary knob/Enter button to select source
Un1S
(Voltage
source from non-isolated connector).
3. Press the Rotary knob/Enter button to commit the selected mode.
LE C
is displayed on the output voltage display.
4. Select the input source level using the Rotary knob/Enter button.
5. Once the desired level has been selected, press the Rotary knob/Enter
button to commit the change.
The display will return to the normal display.
Deactivating APG Current Mode
To deactivate the APG current mode:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
2.
CAPr
is displayed on the output voltage display.
3. Turn the Rotary knob/Enter button to select
nonE
.
4. Press the Rotary knob/Enter button to commit the change.
The display will return to the normal display.
Query for Analog Current Input Level
To query for analog current input level from non-isolated input:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr
is displayed on the output voltage display.
2. If the mode is currently
nonE
then select any other APG mode. If this
step is done, be sure to restore the APG setting to
nonE
when the
operation is complete. Otherwise, do not rotate the Rotary knob/Enter
button to keep the current APG mode.
Содержание XTR 100-17
Страница 2: ......
Страница 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 48: ...2 18...
Страница 86: ...Local Operation 3 38 975 0200 01 01 Figure 3 4 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 122: ...4 30...
Страница 208: ...6 20...
Страница 244: ...A 36...
Страница 250: ...B 6...
Страница 264: ...C 14...
Страница 272: ...IX 4...
Страница 273: ......