Calibration and Troubleshooting
6-2
975-0200-01-01
Introduction
The calibration of the power supply is software dependent, and there are
no potentiometers to adjust.
Calibration is performed via SCPI commands.
The following items need to be calibrated:
•
Programmed voltage
•
Voltage readback
•
Programmed current
•
Current readback
•
Non-isolated voltage monitoring
•
Isolated voltage monitoring
•
Non-isolated current monitoring
•
Isolated current monitoring
•
Isolated source for analog programming
•
Non-isolated source for analog programming
•
Over voltage protection system
Main Voltage and Current Calibration Principle
Understanding the Problem
Figure 6-1 illustrates two sources of analog programming error: gain error
and offset error. Gain error is the departure from the ideal slope of the
measured versus programmed line. Offset error is the magnitude of the
measured value when the programmed value is zero.
For best results, the next recommended action is gain calibration for 90%
of maximum value and offset calibration for 10% of maximum value.
Perform gain calibration again.
CAUTION
Calibration procedures should only be performed by qualified user.
Содержание XTR 100-17
Страница 2: ......
Страница 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 48: ...2 18...
Страница 86: ...Local Operation 3 38 975 0200 01 01 Figure 3 4 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 122: ...4 30...
Страница 208: ...6 20...
Страница 244: ...A 36...
Страница 250: ...B 6...
Страница 264: ...C 14...
Страница 272: ...IX 4...
Страница 273: ......