10. Troubleshooting
Service to be completed by qualified repair personnel only!
10.1 Error Codes
If the X-Cite
®
exacte internal
monitoring systems identify a problem, an error
code will be generated. Definitions and recommended actions are in the
following table.
Code
Error Description
Recommended Action
E1
Lamp temperature is too high
Check ventilation filters and outlets to ensure
that no blockage exists. Restart unit and if
problem persists replace lamp. If problem
continues
. See also
section 10.7
E2
Internal communication failure
Restart unit, if problem persists contact
E3
Iris failed to go to home position
Restart unit, if problem persists contact
E4
Internal hardware failure
Restart unit, if problem persists contact
E5
Failure to communicate with
Intelli-Lamp
Check Intelli-Lamp connection and restart unit.
If problem persists try another lamp, if problem
still persist contact
E6
Lamp self-extinguished
Let system cool and restart lamp. If it goes out
again, replace lamp. If new lamp continues to
self-extinguish, contact
E7
Internal system error
Restart unit, if problem persists contact
bulb
Lamp Error
Lamp installed incorrectly /
Lamp did not strike.
cool
Lamp is too hot to strike
The lamp will automatically strike when it has
cooled to the optimum striking temperature
Alternating:
old / bulb
Old Lamp
The lamp has accumulated over 2500 hours.
Lamp may be near end of life.
Alternating:
end / bulb
New Lamp Required
The lamp has reached end of life. The lamp
will not strike.
LOC
Front Panel Locked
All front panel buttons have been locked to
prevent settings from being changed.
SFI
Shutter Failure
The shutter has failed to return to home
position. Unit should be restarted. If the error
repeats, contact