47
X-Cite
®
exacte
User Guide
035-00283 rev 3
11.6 Replacing the Air Filter
11.6.1 The external air filter is located under the front face plate of the X-Cite
®
exacte
.
11.6.2 Turn off the main POWER switch and remove the AC POWER cord from the unit.
11.6.3 Gently slide and pull out the filter.
11.6.4 Push in the replacement filter so that it sits flat in place
11.7 Replacing the External Fuses
11.7.1 The external (main) fuses are located in the fuse drawer which is located in the
AC inlet module on the rear panel.
11.7.2 Turn off the main POWER switch and remove the AC POWER cord from the unit.
11.7.3 Gently pull out the drawer with the aid of a flat-head screwdriver.
11.7.4 Carefully lever one end of each fuse up from its retaining clip with a small flat-
head screwdriver and lift it out.
11.7.5 Replace the blown fuse(s) only with the same type and rating (F5A, 250V). The
rear compartment must contain two active fuses.
11.7.6 Close the fuse drawer.
11.7.7 Reconnect the AC POWER cord.