31
X-Cite
®
exacte
User Guide
035-00283 rev 3
9.4 Icons and Functions:
ICON
NAME
DESCRIPTION
Alarm
Indicates an alarm state
Click to silence audible alarm
Red = Alarm on
Temperature
Indicates lamp temperature status
Yellow = Lamp running hot
Red = Lamp off, excessive heat
Calibration
Indicates calibration status
Green = Calibrated
Grey = not Calibrated
Lamp Hours
Indicates hours logged on lamp
green<2500 hours, yellow = >2500 hours, make
sure there is a spare lamp on hand
red = >4000 hours, lamp is off due to age.
Light Guide
Indicates light guide position
Green = Light guide present
Red = Light guide not present
Shutter
Click to open/close shutter
[note: lock front panel to use, see note #2
below]
Intensity
Left mouse click = increase by 1% increment.
Right mouse click = decrease by 1% increment.
Left double click mouse = dialog box enabling
data input of desired intensity or power.
Closed-Loop
Feedback
Engage/disengage CLF mode
Grey = OFF
Green = ON, if clicked ON intensity will freeze,
click again to turn off.
Lock/unlock
Lock/unlock the front panel controls on the unit
Green = front panel of X-Cite
®
exacte
locked
Grey = front panel of X-Cite
®
exacte
unlocked,
Shutter will not activate from the control panel if
the front panel is not locked.
Lamp
Click to turn lamp on/off, color indicates status
Green = ON, click to turn lamp on
Grey = OFF, click to turn lamp off
Yellow = Warming Up, lamp is on and is
warming up.
Record
Record CLF data on graph
[% variance from set point]
Pause
Pause recording data
Stop
Stop recording data
Plot/Event
Left-click on graph area to log an event
Notes:
1. Icon background colour code:
Green icons mean status is OK and/or the function is engaged
Yellow icons mean a warning, or that the system is getting ready
Red icons mean there is an alarm condition
Содержание Exacte