2. Shutter control with GUI vs. front panel buttons: The internal shutter can be
operated using either method, however for the unit to accurately monitor
shutter status only one method should be used at a time. If the front panel is
“unlocked” the unit assumes the front panel shutter button will be used and
the shutter icon on the GUI is disabled. I
f the front panel is “locked” the
system assumes that the GUI control will be used and the front panel shutter
button is disabled.
3. When icons are
ALL
grey, the X-Cite
®
exacte
is either disconnected or not
communicating properly with the computer
4. Side Panel Tabs shown greyed-out in Figure 8 GUI Control Panel will not be
operational unless the X-Cite
®
Optical Power Measurement System is
connected. Refer to X-Cite
®
Optical Power Measurement System User Guide
035-00390R for instructions.
5. GUI Control Panel is collapsed in view below.