23
X-Cite
®
exacte
User Guide
035-00283 rev 3
maintain constant intensity, CLF will not be able to compensate for further drops
in power.
Warning signals while CLF is engaged:
1. If the iris reaches 95% open or more, the
CLF
LED will start to blink.
2. If the iris reaches 100% open, the
CLF
LED will blink and alarm will beep until
CLF is disengaged.
NOTE:
If setting CLF while in absolute power mode (display in watts), check the
actual iris setting first to be sure that there is room for output adjustments
to be made.
What will the display show when CLF is engaged?
In Relative Mode: Actual iris position (i.e. value will update as iris is adjusted in
CLF)
In Absolute Mode: Output power (i.e. value will not change, since this is the set
point)
NOTE:
The intensity setting cannot be changed while CLF is engaged, however,
the
MODE
button can be used as usual to toggle between display modes
to view hours, iris position and power set point.
Содержание Exacte