1
X-Cite
®
exacte
User Guide
035-00283 rev 3
1. X-Cite
®
exacte
Message Reference
Display
Message
Definition
Description
XXX
Intensity Setting (%)
Displays the iris setting, which gives relative intensity as
percentage of the lamp’s maximum output
XX.XX
Power Setting (watts)
Displays the power setting, which gives absolute
intensity in terms of watts
uXXX
Power Setting (microwatts)
Displays the power setting, which gives absolute
intensity in terms of microwatts
X.XXX
or
XXX.X
Power Setting (milliwatts)
Displays the power setting, which gives absolute
intensity in terms of milliwatts
XXXX.
Lamp Hours
Displays the “hours of use” accumulated by the lamp
Flashing:
“
XXXX
”
Warming Up
The lamp is warming up. Warm up time is
approximately 4 minutes ref
section 6.1
bulb
Lamp Error
Lamp installed incorrectly /
Lamp did not strike. See
section 10.3
cool
Lamp is too hot to strike
The lamp will automatically strike when it has cooled to
the optimum striking temperature. See
section 10.3
Alternating:
old / bulb
Old Lamp
The lamp has accumulated over 2500 hours. Lamp may
be near end of life.
Alternating:
end / bulb
New Lamp Required
The lamp has reached end of life. The lamp will not
strike.
LOC
Front Panel Locked
All front panel buttons have been locked to prevent
settings from being changed.
SFI
Shutter Failure
The shutter has failed to return to home position. Unit
should be restarted. If the error repeats,
E1
Lamp temperature is too high
Check ventilation filters and outlets to ensure that no
blockage exists. Restart unit and if problem persists
replace lamp. If problem continues
E2
Internal communication failure
Restart unit, if problem persists
E3
Iris failed to go to home position
Restart unit, if problem persists
E4
Internal hardware failure
Restart unit, if problem persists
E5
Failure to communicate with Intelli-
Lamp
®
Check Intelli-Lamp
®
connection and restart unit. If
problem persists try another lamp, if problem still persist
E6
Lamp self extinguished
Let system cool and restart lamp. If it goes out again,
replace lamp. If new lamp continues to self-extinguish,
E7
Internal system error
Restart unit, if problem persists
See Section 10 for troubleshooting.
Содержание Exacte
Страница 1: ...X Cite exacte User Guide...