24
X-Cite
®
exacte
User Guide
035-00283 rev 3
7. Calibration/Radiometry (optional)
All X-Cite
®
exacte
units are capable of being calibrated with an X-Cite
®
Power Meter
XR2100 or X-Cite
®
Radiometer XR2000 (optional accessories). Once calibrated, the
lamp output can be set in absolute units (watts), for easily repeatable illumination levels.
For convenience, the basic instructions for using the X-Cite
®
Power Meter XR2100 or X-
Cite
®
Radiometer XR2000 with the X-Cite
®
exacte
are provided here. For full instructions
and safety information, refer to the complete
Optical Power Measurement System User’s
Guide (035-00390R) or Radiometer XR2000 User Guide (035-00285).
Important Buttons and Settings (underlined settings are recommended for use in
microscopy):
RELATIVE/ABSOLUTE
– toggles between readings in % (relative) and watts
(absolute).
POWER/IRRAD
– toggles between readings in W (power) and W/cm
2
(irradiance); scale will adjust between W and mW automatically (XR2000 only)
X-CITE
®
CAL
– initiates calibration sequence
7.1 X-Cite
®
exacte
Power Reading using the XR2100 Power Meter or
XR2000;
1.
Set the X-Cite
®
exacte
to the desired intensity.
2.
Close X-Cite
®
exacte
shutter using
SHUTTER
button.
3.
Remove light guide from microscope adapter
4.
Snap the 3mm (red) light guide adapter into the top of XR2100 (if not already in
place).
5.
Loosen thumb screw on red light guide adapter, fully insert and fasten light guide
into holder.
6.
Turn XR2100 on using the
ON
button
7.
Open X-Cite
®
exacte
shutter, the power reading will show on the display.