4.5.3 Resetting to default settings (Reset)
Default settings
In a reset, the following functions are reset to the default set-
tings (initialized):
To reset the settings proceed as follows:
The settings have been reset to the default condition. The
instrument continues to operate in the normal measuring
mode. The date and time do not have to be set again.
Note
After a reset, the instrument works temporarily with a factory
calibration. Recalibrate the instrument in order to measure
with the optimum accuracy.
Setting point
Default setting
Unit
NTU
Ratio method
Off
Resolution
0.01 NTU
Code function
Off
Periodical measured value output
On
Output interval
5 seconds
Baud rate
9600 baud
Calibration interval
30 days
1 Press the
4
key and hold it down.
2 Simultaneously press the
8
key. All the display ele-
ments appear (for up to 2 minutes). Subsequently the
instrument switches itself off.
3 Switch the instrument on again.