Technical data
Turb 555 / Turb 555 IR
72
ba41113e07
04/2009
Measuring ranges
Resolution
Accuracy
(± 1 digit)
Reproducibility
± 1 % of the measured value or ± 0.01 NTU
Response time
Less than 6 seconds
Calibration
Automatic 1 - 5 point calibration
Recorder
output
Bi-directional RS232 output
Unit
Turb 555
Turb 555 IR
NTU
0 - 10 000
0 - 10 000
FNU
---
0 - 10 000
EBC
0 - 2 450
0 - 2 450
NEPHELO
0 - 67 000
---
FAU
---
0 - 10 000
In the range
0.0001 - 9.9999 NTU max 0.0001 NTU
In the range
10.000 - 99.999 NTU max 0.001 NTU
In the range
100.00 - 999.99 NTU max 0.01 NTU
In the range
1000.0 - 9999.9 NTU max 0.1 NTU
In the range
0 - 1 000 NTU
± 2 % of the measured value +
0.01 NTU
In the range
1 000 - 4 000 NTU
± 5 % of the measured value
In the range
4 000 - 10 000 NTU
± 10 % of the measured value