4.2.2 Single measurements
Preparatory activi-
ties
Perform the following preparatory activities when you want
to measure (see section 3.3 S
WITCHING
ON
THE
MEASURING
INSTRUMENT
):
Note
Incorrect calibration of the turbidimeter will result in incorrect
measured values.
Carry out the calibration regularly at the specified fixed inter-
vals.
Note
The outside of the cuvette always has to be clean, dry, and
free of fingerprints. Clean the cuvette before starting to mea-
sure (see section 5.2.2 C
LEANING
THE
CUVETTES
). Only hold
the cuvettes by the top or by the black light protection cap.
Measuring
How to measure the turbidity of a sample:
1 Switch on the measuring instrument.
2 Wait until the warm-up period is over.
1 Rinse out a clean cuvette with the sample to be mea-
sured:
Pour approximately 20 ml sample into the cuvette.
Close the cuvette and rotate it several times before
throwing the sample away.
2 Repeat the rinsing procedure twice more.
3 Fill the cuvette with the sample to be measured (ap-
prox. 30 ml). Close the cuvette with the black light
protection cap.
4 Make sure that the outside of the cuvette is clean,
dry, and free of fingerprints.
Содержание Turb 555
Страница 1: ...ba41113e07 04 2009 Turb 555 Turb 555 IR Operating Manual Turbidity measuring instrument...
Страница 12: ...Safety Turb 555 Turb 555IR 12 ba41113e07 04 2009...
Страница 16: ...Commissioning Turb 555 Turb 555IR 16 ba41113e07 04 2009...
Страница 62: ...Operation Turb 555 Turb 555IR 62 ba41113e07 04 2009...
Страница 66: ...Maintenance cleaning disposal Turb 555 Turb 555IR 66 ba41113e07 04 2009...
Страница 74: ...Technical data Turb 555 Turb 555IR 74 ba41113e07 04 2009...
Страница 81: ......