Turb 555 / Turb 555 IR
Commissioning
15
ba41113e07
04/2009
3.3
Switching on the measuring instrument
Sample display
Warm-up period
For routine measurements without special accuracy require-
ments, a warm-up period of 5 minutes is sufficient. For high
precision measurements, the warm-up period should be 30
minutes minimum. This is also valid when you want to cali-
brate your instrument. Only then we can guarantee that the
instrument measures with the accuracy specified in
chapter 7 T
ECHNICAL
DATA
.
Note
Even if the instrument is not measuring, always leave a cu-
vette plugged in the cuvette shaft so that the cuvette shaft is
protected against dust.
1 Place the measuring instrument on a flat surface and
protect it from intense light and heat.
2 Press the
0
key.
The measuring instrument automatically switches to
the measuring mode: A measured value and the
a
display appear.
3 Set the date and time, if necessary (see section
4.5 S
ETTING
UP
THE
INSTRUMENT
).
4 Wait until the warm-up period is over.