Turb 555 / Turb 555 IR
Maintenance, cleaning, disposal
63
ba41113e07
04/2009
5
Maintenance, cleaning, disposal
5.1
Replacing the lamp module
The lamp module can easily be replaced if it is defective.
You will find the order number for the spare lamp module in
chapter 8 A
CCESSORIES
, O
PTIONS
. Proceed as follows to re-
place the module:
1 Switch the measuring instrument off and disconnect
it from the mains.
2 Take out the lamp module (3): Press the two grips
gently together and carefully pull out the lamp mod-
ule until the plug-in connection (2) is visible.
3 Pull the plug-in connection (2) apart and, thus, re-
move the lamp module. Do not pull on the wires while
doing this.
4 Plug in the new lamp module to the plug-in connec-
tion of the instrument. Do not touch the lamp with the
fingers.
5 Stow the wires in the opening (1) of the measuring in-
strument again. Make sure that the wires are not
placed in front of the lamp and that they are not in the
way of the lamp module.
Содержание Turb 555
Страница 1: ...ba41113e07 04 2009 Turb 555 Turb 555 IR Operating Manual Turbidity measuring instrument...
Страница 12: ...Safety Turb 555 Turb 555IR 12 ba41113e07 04 2009...
Страница 16: ...Commissioning Turb 555 Turb 555IR 16 ba41113e07 04 2009...
Страница 62: ...Operation Turb 555 Turb 555IR 62 ba41113e07 04 2009...
Страница 66: ...Maintenance cleaning disposal Turb 555 Turb 555IR 66 ba41113e07 04 2009...
Страница 74: ...Technical data Turb 555 Turb 555IR 74 ba41113e07 04 2009...
Страница 81: ......