Maintenance
NOx-N
3 - 42
ba43110e05
01/2011
3.11
Switching Auto Flow on/off
The AutoFlow program is used to measure the current fluid levels of
sample and reagents.
z
Error messages are generated.
z
The fluid levels of the reagents are determined.
Fluid level measurement
If the
Fluid level measurement
parameter is switched on, the current
fluid levels in the containers are continuously determined by pressure
measurements. If the fluid level sinks below 4 cm, an instruction is
issued to fill the container, e.g. "Replenish dist. H2O".
Overpressure
measurement
If the
Overpressure measurement
parameter is switched on, the
sample flow is monitored.
If there is a lack of sample, the analyzer module signals "No sample"
until sample is available again or until 15 minutes have passed. During
this time, the peristaltic pump operates at low speed. The pressure
measurement is performed continuously during this time so that a
measuring cycle starts immediately when sample is available again.
If no sample is available after 15 minutes, the analyzer module signals
"NO SAMPLE".
Relationship
Parameter/messages
If the relevant parameter is switched on, messages are output
according to the following table:
Note
Further information on how the pressure sensors function is given in
the appendix, section 6.2 P
on page
Parameter
Values
Messages
Fluid level
measurement
On, Off
Replenish dist. H2O
Dist. H2O blocked
Dist . H2O empty
Replenish cleaning solution
Cleaning solution blocked
Cleaning solution empty
No sample
Test sample blocked
Overpressure
measurement
Off, On
Outlet clogged