Operation
NOx-N
2 - 24
ba43110e05
01/2011
Parameter
Values
-no
designation-
Display of whether AutoAdapt is active (can be
changed in the
Measuring Interval
menu).
DC/time
0.1 - 10.0 mg/L/Measurement (NOx-N)
Minimum
interval
Continuous, 5, 10, 15, 20, 25,
30 min
The minimum
interval must be
selected to be
lower than the
maximum
interval!
Maximum
interval
5, 10, 15, 20, 25, 30 min
1
Measuring mode
Select TresCon NO
X
-N with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
AutoAdapt
menu item with
68
and confirm
with
3
. The prompt
Presentation of data/Change
appears.
4
Select the
Change
mode with
68
and confirm with
3
.
5
Enter the PIN digit-by-digit with
68
. Change to the next
digit with
57
and confirm with
3
.
The analyzer switches into the stop mode.
6
Select the parameters to be changed with
68
and
switch into input mode with
3
.
Changing the
AutoAdapt parameter
Содержание TresCon NOx-N