NOx-N
Maintenance
3 - 5
ba43110e05
01/2011
Caution
Only use distilled water without any contamination (softener orsimilar)!
Dist. H2O is also used for zero point calibration. Contaminated dist.
H2O causes incorrect measurement results.
1
Measuring mode
Select the NOx-N measuring module
WITH
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Fill system
menu item with
68
and confirm
with
3
. A checklist appears on the display.
5
Place a full container on the tray.
6
Connect the consumables. Follow the color coding!
red
Dist. H2O
blue
Cleaning solution
7
Confirm the prompts each time with
3
after checking. After
the last confirmation, the procedure begins immediately and
the remaining duration appears on the display.
Starting the procedure
Color coding
of the consumables