NOx-N
Maintenance
3 - 27
ba43110e05
01/2011
Replacing the
tubing system
5
Measuring mode
Select the NO
X
-N analyzer module with
68
.
6
Call up the
Parameter NOx-N
menu with
3
. The main menu
appears on the display.
7
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
8
Select the
Replace tubing system
menu item with
68
and confirm with
3
. The 1st part of the checklist appears.
9
Confirm each of the requests of the 1st part of the checklist
with
3
after checking them. After the last confirmation, the
procedure begins immediately. The duration appears on the
display.
10
After completion of the 1st part of the routine, the 2nd part of
the checklist appears on the display.