10
Swing out the pump cassette. To do so, open the gripping
lever. Subsequently, unlock the pump cassette by pressing on
the gripping lever and at the same time swing out the pump
cassette.
11
Displace the pump tubing. Insert the pump tube in the pump
cassette so that the first (used) section is on the pressure side.
12
Click the cassette into place in the peristaltic pump and close
the gripping lever.
Standard/sample pump tubing:
blue/blue/blue
Содержание TresCon NOx-N