Maintenance
NOx-N
3 - 26
ba43110e05
01/2011
The service menu "guides" you through the necessary steps. The
service routine consists of two parts.
Duration
z
Part 1: approx. 1
½
minutes
z
Part 2: approx. 1
½
minutes
This service resets the timer of the interval,
Replacing the tubing
system
. An AutoCal is triggered when the subsequent measuring
procedure starts. You cannot quit the procedure.
Caution
The pumps empty the system via the suction tubes. If the containers
with the consumables are not removed, the distilled water can become
unusable! You cannot quit the procedure.
Caution
After step 9 of the following instructions, you can no longer quit the
procedure!
Preparation
Step
Instrument status/action
1
Flush back sample
Replace tubing system. During this operation, the TresCon NOx-N
remains in the Wait mode. After confirmation of the tasks by pressing
a key, the sequence of operations is continued.
2
Rinse sample
Part 1
Emptying the system
Part 2
Filling the system
1
Pull the sample tubing out of the overflow vessel.
2
Unscrew all tubings from the consumable containers.
3
Place a suitable collecting container (recommended volume
0.5 l) on the tray of reagents.
4
Insert all ends of the tubes into the collecting container.