NOx-N
Appendix
6 - 5
ba43110e05
01/2011
TXRCINT
Read the following settings:
z
AutoCal: Time, interval
z
AutoClean: Time, interval
Instrument response:
TXRCONZ
Read the current concentration.
Instrument response:
TXRGPAR
Read the instrument parameters.
Instrument response:
--------------------------------------------
NOx-N Module 1
22.04.03 15:30:37
--------------------------------------------
AutoCal: 07:00, 24 h
AutoClean: 07:00, 24 h
--------------------------------------------
TresCon>
--------------------------------------------
NOx-N Module 1
22.04.03 15:32:01
--------------------------------------------
1.22 mg/l NOx-N
--------------------------------------------
TresCon>
-----------------------------------------------------------
Selected parameters
NOx-N Module 1
22.04.03 15:32:42
-----------------------------------------------------------
Display : NOx mg/l
AutoCal time : 07:00 h
AutoCal interval : 24 h
AutoClean time : 07:00 h
AutoClean interval : 24 h
Measuring interval : Continuous
Manual correction : Off
Attenuation : 2
AutoFlow
Fluid level measurement : On
Overpressure measurement : On
Dilution: : None
-----------------------------------------------------------
TresCon>