Appendix
NOx-N
6 - 2
ba43110e05
01/2011
6.2
Pressure sensors and messages
Note
During pressure measurement, the valves are immediately in the mode
before the pressure measurement and the pump is off.
Display on the
screen
Type of error
Pressure sensor
Fill dist. H2O
Dist. H2O must be
replenished
P1
-60 cm < h1 < 4 cm
Dist. H2O clogged
Dist. H2O clogged
P1
h1 < -60 cm
Dist. H2O empty
Dist. H2O empty
P1
h1 > 50 cm
Display on the
screen
Type of error
Pressure sensor
Replenish cleaning
solution
Cleaning solution must
be filled
P1
-60 cm < h1 < 4 cm
Cleaning solution
clogged
Cleaning solution
clogged
P1
h1 < -60 cm
Cleaning solution
empty
Cleaning solution
empty
P1
h1 > 50 cm
Display on the
screen
Type of error
Pressure sensor
Sample inlet clogged
Sample inlet clogged
P1
h1 < -60 cm
Sample missing
Sample missing
P3
P3 < 10 mbar
Outlet clogged
Outlet clogged
P3
P3 > 150 mbar
Dist. H2O
Cleaning solution
Sample inlet
Содержание TresCon NOx-N
Страница 20: ...Overview NOx N 1 16 ba43110e05 01 2011 ...
Страница 92: ...Maintenance NOx N 3 44 ba43110e05 01 2011 ...
Страница 98: ...What to do if NOx N 4 6 ba43110e05 01 2011 ...
Страница 108: ...Appendix NOx N 6 8 ba43110e05 01 2011 ...
Страница 111: ......