NOx-N
Overview
1 - 13
ba43110e05
01/2011
1.2.4
Initial commissioning
When the TresCon NOx-N is installed and connected, the initial
commissioning can start.
Note
Read the B
ASIC
OPERATION
section in the T
ERMINAL
volume of this
operating manual if you are not yet familiar with the operation of the
TresCon.
1
Switch on the TresCon at the power switch (mounting stand).
The analyzer modules warm up. The TresCon NOx-N is ready
for operation after it reaches the operating temperature.
2
Measuring mode
Select TresCon NO
X
-N with
68
.
3
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
4
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
5
Perform the
Fill system
service according to section 3.2, page
3-4 of this operating manual.
Switching on the
TresCon
Filling the system