Use
17.10.2019
59 / 96
Advanced programming
a
b
c
d
a) In test mode,
select
Precise
to enable leak rate measurement.
ð
In addition to the result of the leak test (OK or not OK), the
leak rate is displayed.
ð
Displaying a leak rate in "Precise" test mode requires more
time for the measurement.
b) If necessary, set a test time for each test.
ð
Entering a manually set measuring time may mean that the
measurement algorithm cannot identify enough measuring
points to produce a reliable measurement result. The user
must therefore ensure that the selected measuring time
enables the required test accuracy to be met.
ð
A number field appears for entering the times.
c) If necessary, set a maximum leak rate test in µg/min for each test.
ð
A number field appears for entering the leak rates.
d) If necessary, set the minimum and maximum deformation limit in
µm for each test.
ð
A number field appears for entering the minimum and
maximum deformation.
e) Press
Basic settings
to enter the basic programming screen or
Save
to save the program.
5.5.4
Editing a program
a) Select the user program to be edited.
ð
See
.
b
b) Press
to display the user program configuration window.
c) Change and save the desired settings.
ð
See the
Basic programming
and
Advanced programming