Use
17.10.2019
57 / 96
Abbreviation
Program name
Pressure values
(bar)
Limit values
(µg/min)
Watch size
(mm)
CL
Calibrated Leak
2.0
50
20–40
P01
Safe test
-0.2/2.0
200/200/50
20–40
P02
Standard Men
-0.4/3.0
200/50
20–40
P03
Ladies
-0.4/2.0
200/50
< 20
P04
Jewelery
-0.2/1.0
200/50
< 20
P05
Diver 100 m
-0.7/1.0/10.0
200/200/50
20–40
Info
The 50 µg/min limit is only applied at the highest test
pressure of a program, as per ISO 22810. The previous test
pressures are tested for the preliminary test with a limit of
200 µg/min.
5.5.3
Creating a program
Creating a program
a) Press
to display the program menu options.
b) Press
to open the program menu options.