![Witschi ProofMaster Скачать руководство пользователя страница 64](http://html.mh-extra.com/html/witschi/proofmaster/proofmaster_operating-manual_990032064.webp)
Maintenance
64 / 96
17.10.2019
6.3
Cleaning the seal (O ring)
ü
The ProofMaster is switched off and the bell cover is open.
1
a) Clean the seal
1
with a microfibre cloth.
6.4
Replacing the seal (O ring)
ü
The ProofMaster is switched off and the bell cover is open.
2
1
a) Lift the seal
2
with the push button
1
and remove by hand.
b) Remove dust particles and other deposits from the seal groove
2
.
c) Insert a new seal
2
.
6.5
Replacing the deformation sensor
2
1
ü
The ProofMaster is switched off and the bell cover is open.
a) Unscrew screw
1
with a T10 Torx screwdriver.
b) Remove the deformation sensor
2
.
c) Hook in the new deformation sensor
2
.
d) Tighten screw
1
.