Operation Manual CoatMaster
47/83
When the checkbox Expert is activated two additional tabs for further settings are available.
Evaluation:
Allow variation of substrate properties
: This checkbox should be activated if the calibration and the
measurement are carried out on different substrate types.
NOTICE
However, for optimum results we recommend to carry out the calibration and the measurement un-‐
der as similar conditions as possible.
Force correlation with reference
: This checkbox should be activated when there is a bad thermal
contrast between coating and substrate, i.e. they behave nearly identically. In this case the meas-‐
urement values can widely vary, this setting has a stabilizing effect.