Operation Manual CoatMaster
42/83
NOTICE
A submenu is available when right-‐clicking the chart of the measurement points, enabling to save and
print the chart, zoom in and out and defining an auto range.
Here, the most important measurement parameters, a straight line through the measurement points
and possible error messages are displayed.
Mean deviation:
> 25 %: at least one of the measurement methods does not work, change the reference method
between 10 and 25 %: measurements not optimal, repeat measurements
< 10 %: value is okay
Quality index:
> 94 %: very good
Between 90 and 94%: Repeat blank measurement (which was done prior to the first putting into
operation)
< 90 %: not okay, a reason might be that wrong materials were chosen for substate or coating or that
the materials show an untypical behavior. In this case, choose the entry “unknown” for substrate or
coating material