8. Remote Operation
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
71
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
Item
Command or Query
Response/Function
UNIT
46.
:NAME<n>?
Returns the name of the unit with index n.
47.
:FACTOR<n>?
Returns the conversion factor to the current unit
with index n.
48.
:PRESsure<n>
Sets the pressure units.
49.
:PRESsure?
Returns the pressure units.
OUTPut
50.
:STATeON/OFF/1/0
ON or 1 = Control OFF or 0 = Measure
51.
:STATe?
Returns 0 for measure 1 for control.
52.
:MODE MEASure/CONTrol/VENT
Sets the mode indicated.
53.
:MODE?
Returns the mode string.
54.
:STABle?
Returns 1 if stable 0 if not.
55.
:AUTOvent ON/OFF/1/0
Puts the unit in the Vent mode if true
56.
:AUTOvent?
Returns the state of the Vent mode.
57.
:AUTORange ON/OFF/1/0
Turns autorange on or off
58.
:AUTORange?
Returns “ON” or “OFF”
[SOURce]
:PRESsure
[:LEVel]
[:IMMediate]
59.
[:AMPLitude] <n>
Sets the setpoint.
60.
[:AMPLitude]?
Returns the setpoint.
61.
:SLEW <n>
Sets the slew rate setpoint.
62.
:SLEW?
Returns the slew rate setpoint.
63.
:TOLerance <n>
Sets the stable window.
64.
:TOLerance?
Returns the stable window.
CALCulate
:LIMit
65.
:LOWer<n>
Sets minimum control limit.
66.
:LOWer?
Returns minimum control limit.
67.
:UPPer<n>
Sets maximum control limit.
68.
:UPPer?
Returns maximum control limit.
:SYSTem
69.
:DETECT
SLOW/FAST/CANCEL
Control autotune
(NOT USED NOW).
70.
:DETECT?
Returns state of control autotune.