9. Calibration
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
77
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
of the internal transducer for each pressure point can be seen directly in the “Reading” window.
Pressing the corresponding point under the “Actual” column will accept that re
ading into that point.
Subsequent pressure points generated by the pressure standard are accepted and recorded in the
same way. After all points are entered, press the [Save] button to linearize the internal sensor based
on the “Live Cal” data.
Figure 9-6 Linearization - Live cal
At any time the [Factory Cal] button can be pressed to revert back to the factory calibration.