7. Operation
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
51
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
USB Device Setting
The USB Device jack is a standard USB Type B receptacle interface used for remote communication.
The USB Driver can be downloaded at:
http://www.mensor.com/download_software_instrument_en_um.wika
7.3.5 Applications Tab
Press the [Applications] tab in the Setup [
] screen and a screen will appear containing various
labeled icons that when pressed access other screens that provide the following functions:
Calibration of internal transducers
Sequence program configuration
Selection of favorites viewable in the main operation screen
A troubleshooting screen that displays errors
The advanced screen that is used for tuning the regulator and viewing sensor linearization
An administration screen used to manage passwords
Figure 7-43 Applications Tab - Locked
Passwords
All the calibration screens
are password protected with the “Calibrate Password”
[1 Point Calibration]
[2 Point Calibration]
[Linearize]
[Head Correction]
The “Service
Password” allows access to:
[Advanced]
[Admin]
Figure 7-44 Applications tab - Unlocked
shows the [Applications] screen when it is unlocked.
INFORMATION
The Default Passwords sent with the instrument are as follows:
Calibrate Password: 123456
Service Password: 987654
Both passwords can be changed and saved in the [Admin] page within the [Setup] [
]
/ [Applications] section.
If the passwords are changed then forgotten contact Mensor or your local Mensor
service center for instructions to access the protected areas and to reset the passwords.