8. Remote Operation
60
WIKA Operating Instructions High Pressure Controller, Model CPC8000-H
GB
1
4
2
3
5
9
9
3
.0
0
/0
6
/2
0
1
7
/GB
Item Command or
Query
Data
Response/Function
28.
CRATE?
<sp>CCCCCC<cr><lf>
Returns the control rate
–
CCCC is variable in length
and corresponds to the
parameters for the
CRATE command.
29.
CTYPE?
Returns the type of
regulator.
30.
DECPT
4, 5, 6 or 7
Sets the number of
decimal points.
31.
DECPT?
<sp>n<cr><lf>
Returns the number of
decimal points. (See:
Resolution)
32.
DEFAULT
none
Sets the default values.
33.
DHCP
ON or OFF
Reserved for DHCP
setup.
34.
DHCP?
<sp>(YES or NO)<cr><lf>
Reserved for DHCP
setup.
35.
DIO
2 or 0
2 turns on the first digital
output pin, 0 turns it off.
36.
DIO?
<sp>n<cr><lf>
Returns status of the first
input and output pin. Bit0
= input’s status, bit1 =
output’s status.
37.
DOUTFUNC
<n><sp><CCCCCC><cr><lf>
Sets the function for
output pin <n> to NONE,
STABLE, PUMP,
MEASURE, CONTROL,
or VENT.
38.
DOUTFUNC?
<sp><CCCCCC><cr><lf>
Returns the function for
output pin <n>.
39.
DINFUNC
<n><sp><CCCCCC>
Sets the function for input
pin <n> to NONE,
MEASURE, CONTROL,
VENT, KEYLOCK, or
START.
40.
DINFUNC?
<sp><CCCCCC><cr><lf>
Returns the function for
output pin <n>.
41.
DOUTSTATE
<n><sp>HIGH/LOW/1/0>
Turns output pin <n> to
high or low. Sets the
function for that pin to
“None”.
42.
DIOSTATE?
<sp><n><cr><lf>
Bit 0-7 is the status of the
input bits. Bit 8-15 are the
staus of the output bits.
Returns an integer
between 0 and 65536.
43.
DOC
mm/dd/yyyy
Sets the date of cal for the
active channel.
Item Command or
Query
Data
Response/Function