79
AECG100 | User Manual
Parameter
Specification
PI
(
AC level/DC level
)
Range
0.5% ~ 15%
Accuracy
x
LED scan rate
rate
1K (Normal mode)
10K (Raw data mode)
Accuracy
±
5μs
PD sample rate
rate
250K (single channel)
Accuracy
±
5μs
PD response time
Rising
1
μ
s Typical
Falling
1
μ
s Typical
SpO
2
(%)
Range
1% ~ 100%
Resolution
1%
Accuracy
91 % to 100%: ±1% +specified
accuracy of the DUT
81 % to 90%: ±2% +specified
accuracy of the DUT
71 % to 80%: ±3% +specified
accuracy of the DUT
Below 70%: with unspecified
accuracy
MCX LED voltage
Amplitude
AC Level x 100
Accuracy
±
5%
§
The data is collected by connecting the spectrometer with 2" integrating sphere and
fastening DUT with the integrating sphere to measure the absolute irradiance value of
DUT LED.
* The PD response time will vary with the light intensity of the DUT.