AECG100 | User Manual
Note:
Standalone (device Mode A/B/C) and PC file can be used in (1) Signal
tab “Recall” and Auto Sequence Mode. Only Waveform Raw can be
played in (3) Player tab.
(6)
Recall:
Load the previously saved test parameters stored in Standalone
(device Mode A/B/C) or as a PC file (*.ppg format).
(7)
Play/Stop:
Play the selected waveform with preset parameters. Once
stopped, play from the beginning.
(8)
Waveform Display Scale:
Adjust the display scale for 2 sec, 5 sec, or 10
sec time frames.
(9)
Inverted/Sync Pulse:
Table 1: Inverted/ Sync Pulse Function Description
Inverted: On
AECG100 outputs PPG waveform upside down.
Inverted: Off
AECG100 output PPG waveform as per the
window display.
Sync Pulse: On
LED behaviors of AECG100 synchronize with
DUT LED.
Sync Pulse: Off
AECG100 LED remains on.
(10)
Sampling:
To sample and measure the DUT LED brightness.
(11)
Signal:
PD-Green/ PD-RED/ PD-Infrared:
It shows the intensity and behaviors of DUT LED sampled by
AECG100
’s PD.
Displayed in
Blue
and
Purple
curves.
Switch-Green/ Switch-RED/ Switch-Infrared
:
It shows the timing AECG100 switches LED on or off. Waveform displays
low means AECG100 turn on LED. Waveform displays high means
AECG100 turns off LED. Displayed in
Blue
and
Red
curves.
PD/ Switch selection is memorized.
(12)
Freeze:
Once checked, the display of PD/ LED switch sampling would
stay as it is.