63
AECG100 | User Manual
Table 3: Inverted/ Sync Pulse Function Description
Inverted: On
AECG100 outputs PPG waveform upside down.
Inverted: Off
AECG100 output PPG waveform as per the window
display.
Sync Pulse: On
LED behaviors of AECG100 synchronize with DUT
LED.
Sync Pulse: Off AECG100 LED remains on.
(10)
Sampling:
Use PD of AECG100 to measure intensities of DUT LED.
(11)
Signal:
PD-RED/ PD-Infrared:
It shows the intensity and behaviors of DUT LED sampled by AECG100’s
PD. Displayed in
Blue
and
Purple
curves.
Switch-RED/ Switch-Infrared
:
It shows the timing AECG100 switches LED on or off. Waveform displays
low means AECG100 turn on LED. Waveform displays high means
AECG100 turns off LED. Displayed in
Blue
and
Red
curves.
PD/ Switch selection is memorized.
(12)
Freeze:
After enabling this function, the display of PD/ LED switch
sampling would stay as it is.
(13)
Max Peak:
Display the peak value of DUT LED, sampled by AECG100
PD.
(14)
Time Axis:
Adjust the time axis of PD sampling and switch.
(15)
Enlarge:
Click to better view the sampling details.
(16)
Load:
To load PPG raw data for each channel (PPG RED/ PPG Infrared).
Select channel first, and then load PPG RED or PPG Infrared raw data in
the
“
File
”
section accordingly. Raw data file is either saved from (1)
S
ignal tab “Waveform Raw” or self
-constructed according to WhaleTeq
rule (click
“
Help
”
for more details). “
Total Length
”
indicates the total
waveform playback time. User has to manually adjust
“
Output Setting
”
to assure AECG100 output signal with suitable parameters.