reposition DUT and repeat Step 2 and Step 3 until
DUT can get the heart rate.
2.4.2
Transmission SpO
2
Testing Device
Figure 14: Transmission SpO
2
Module (PPG-2TF-660)
Step 1:
Clamp the DUT to the index finger simulation on the PPG-2TF-
660, and plug the DB15 connector on the PPG-2TF-660 into the
DB15 port on the AECG100 main console. Plug the USB Type-B
connectors of the USB cables provided to the AUX PWR port and
USB port on the AECG100 main console; at the other end of the
USB cables, plug the USB Type-A connectors into a laptop or a PC.
Switch the power switch located at the lower-left corner on the top
of the AECG100 main console to ‘’On’’ to power on
it.
Note: Align the LED and PD on DUT to the PD and LED on the PPG-
2TF-660 respectively.
Step 2:
Go to the
AECG100 software. Complete the installation and click the AECG
icon to start the test. Select the Auto Test SpO
2
tab and check the
"signal strength" on the software interface.
It shows the optical
signal intensity detected.