AECG100 | User Manual
Upper section:
Figure 55: Auto Test SpO
2
Interface
(1)
BPM:
Adjust for different heart rates. The value is from 10 BPM to 300
BPM.
(2)
SpO
2
:
Adjust for different heart rates. The value is from 1% to 100%.
(3)
Transmission Rate:
The user can choose to simulate 6 different
transmission rates (Light & Thin Finger, Light & Medium Finger, Normal
Medium Finger, Dark Medium Finger, Dark & Thick Finger and Neonatal
Foot).
(4)
Signal Strength:
Detect the signal strength of the DUT to set up a
proper measuring position for test.
(5)
Table:
The dropdown list provides 2 reference tables (Masimo and
Nellcor) for users
to compare. Click the ‘’Table’’
to create
the DUT’s
SpO
2
table for verification. Please refer to the instructions below.
Note:
The Masimo and Nellcor's R curve samples are not provided by their
respective owners. The data are collected through the AECG100 main
console measurement and for reference only.
*The UI is illustration only. For detailed display please check on your
software UI.
Set up the SpO
2
parameter and transmission rate, then
click ‘’
P
lay’’
at the
lower section of the Auto Test SpO
2
tab to send signals to the DUT as figure
56 shows.
2
5
3
4
1