74
AECG100 | User Manual
USB cable, or try turning off and then turning on the AECG100 test
system.
Cautions
(1)
Mode A/B/C is saved in AECG100 main console, while info stored
within Mode A/B/C may correspond to different variable values for
different PPG modules connected. Therefore, when testing
AECG100 main console with different PPG modules, re-saving Mode
A/B/C is required to obtain desired test results.
(2)
When a stable and firm connection between AECG100 main console
and PPG modules is desired (e.g., moving the entire system around),
it is recommended to fasten with a DB15 cable in between. The
cable spec is DB15 male to female.
(3)
It is suggested to warm up AECG100 Test System for at least 30
minutes before testing.
(4)
Before using PPG-related functions, please assure the distance
between the PPG module head and DUT PPG sensor is fixed.
(5)
During testing, it is essential to minimize the ambient light, as it
would influence the test result as well.
(6)
LED luminous decay occurs as time goes on. It is recommended to
send the AECG100 units to WhaleTeq for calibration every year to
replace the LED base if needed (depends on usage frequency) to
ensure LED brightness is qualified. Please contact WhaleTeq Co, Ltd
for more details.
(7)
When upgrading AECG100 firmware, DO NOT disconnect the USB
cable or close the software during the process.
(8)
Warranty voids if QC PASS label is removed or tampered with.
(9)
This is a professional testing instrument instead of a medical device.
It is for testing only and will not be used on human or clinical use.