AECG100 | User Manual
(1)
Electrode:
Select the output electrode to RA/R or LA/L.
(2)
ECG Waveform Parameter:
Set test parameters of voltage and time.
(3)
BPM:
Adjust different heart rates for testing.
(4)
Sync BPM:
BPM values of ECG wave and PPG wave are locked to be
synchronized.
(5)
Load ECG Mode Waveform:
Load and display ECG mode preset
waveform.
(6)
Load PPG Mode Waveform:
Load and display PPG mode preset
waveform.
Not allowed when PPG BPM differs from that of ECG’s.
(7)
PPG Waveform Parameters:
PPG waveform can be adjusted via PI
(Perfusion Index) = AC/DC. Select
“L
ock AC
”
to adjust DC and PI
values. Select
“L
ock DC
”
to adjust AC and PI values.
(8)
Time Difference:
Adjust the time difference between ECG (R wave
peak) and PPG waveform (peak or trough). PTTp is the time to peak
and PTTf is the time to trough.
(9)
Database Player:
In PWTT test mode, AECG100 can download and play
PhysioNet MIMIC database (mimicdb). Furthermore, the WhaleTeq
database includes 30 records of PPG waveforms recorded in clinical
trials to use in PPG test mode and 10 records of PPG and ECG
waveforms recorded in clinical trials to use in PWTT test mode. The
recorded PPG disease waveforms include AF (Atrial Fibrillation), APC
(Atrial Premature Complexes), VPC (Ventricular Premature
Contraction), and first-degree AV block. Additional 5 records of AF
(Atrial Fibrillation) are provided for PWTT test mode. The first record is
available for trial purposes and illustrated in the Player tab from the
lower Section. Please contact WhaleTeq to purchase the completed
database playback function.
(10)
Auto PWTT:
Set time difference, PTTp range, PTTp step, and intervals
for PWTT automation.
(11)
R/IR Selection:
Click to switch between RED and IR LED test
parameters page.