72
AECG100 | User Manual
Figure 86: DC Voltage Verification (300mV)
5.
Connect the multimeter to RA and LA of the AECG100 and measure
the DC voltage. The value should be in 1% of 300 mV. The setup is
shown in the below figure.
Figure 87: Self-Calibration Setup
6.
If the user wants to check the DC voltage other than 300 mV, set
the DC voltage from -1000 mV to +1000 mV with 5% accuracy.
Please look at the figure below for details.
Note: This extra DC voltage can support some other than 300 mV
testing.